the pattern of applied flow to the patient. Square or Decel. are selectable for this
Flow Pat.
parameter.
If "Leak Compensation" is enabled, the ventilator will compensate leaks
Leak Comp.
automatically.
Ability to use the device non-invasively (NIV) and using a mask. This feature is
NIV
defined in P-SIMV and PSV modes.
In volume modes, if patient tries to breathe during inhalation, the device responds
Dual control
to this request by switching on the pressure control. This feature can be selected in
square flow pattern.
The exhalation phase continues until the expiratory flow reaches a certain
Auto Tlow
percentage of the expiratory flow peak, which is equal to the Auto Tlow
adjustment value. This feature is selectable in APRV mode.
5-5

Main Screen

The ventilator main screen consists of different parts (See the figure below).
1. Date, time and battery icon
2. Active ventilation mode
3. Functional alarms (e.g. messages related to activation of maneuvers and apnea mode
inactivity)
ain screen
Figure 5-7 M
32