It should be noted that Back light, Touch Sound and Bed Number items are inactive.

5-10 Graphic Settings

This page is currently inactive.

5-11 Clinical Setup

The following items could be set in this menu:
 Smart nebulizer
 Humidification type
 Compliance compensation
 O2 Sensor
 EXH Flow Sensor
 Smart nebulizer
Activation, deactivation and duration of nebulizer operation can be specified in this section.
By activating the nebulizer from this menu, the nebulizer activates for the time specified by
the user and the related message is displayed on the screen. After this period, it will be
deactivated automatically.
Available options for the duration of nebulizer activity are: 1 min to 480 min. Default setting
is 20 min. Any change in the nebulization time setting will be saved until the next change.
Nebulizer is not activated in inspiratory flow (air and oxygen) less than 11L/min and "Neb.
Not Usable" alarm message will be displayed.
Figure 5-15 Clinical Setup
39