Pause (insp plateau)
Accuracy: ±0.02 sec
Rise time
Exhalation sensitivity
Accuracy: ±12% or ±(3 l/min +4% Actual reading)
wig
Oxygen (FiO2)
Accuracy (Delivery): ± (3 %) full scale
Phigh
Accuracy:±12% or ±(2cmH2O +4% Actual reading)
wig
Plow
Accuracy: ±12% or ±(2cmH2O +4% Actual reading)
wig
Thigh
Accuracy: ±0.02 sec
Tlow
Accuracy: ±0.02 sec
Auto Tlow
Accuracy: ±12% or ±(3 l/min +4% Actual reading)
wig
Nebulizer
Accuracy: ±10 s
Humidification Type
Compliance Compensation
Graph Setting
Trend Data
Trend resolution
Displayed Measurements
Audio/ Settings
Pressure Values
Ppeak (peak pressure during a breath)
Accuracy: ±12% or ±(2cmH2O +4% Actual reading)
wig
PEEP (pressure at end exhalation)
Accuracy: ±12% or ±(2cmH2O +4% Actual reading)
wig
Pmean (averaged mean pressure)
Accuracy: ±12% or ±(2cmH2O +4% Actual reading)
wig
Monitored /
P
Displayed Patient
endInsp
Values
Accuracy: ±12% or ±(2cmH2O +4% Actual reading)
wig
Volume / Flow Values
Vte (expiratory tidal volume)
Accuracy:
For 50- 2600 ml: ±(4 ml + 15% Actual reading)
For 20- 49 ml: ±(7 ml + 15% Actual reading)
Vti
Accuracy:
For 50- 2600 ml: ±(4 ml + 15% Actual reading)
For 20- 49 ml: ±(7 ml + 15% Actual
0 – 75% of Inspiratory time
Fast, Med, Slow
5 – 85 % of peak flow
21 – 100 %
5 – 50 cmH2O
0 – 47 cmH2O
1.0 –29.8 sec
0.2 –22.0 sec
5 – 85 % of peak flow
On or Off;
User selected duration (1– 480
minutes)
None, HME, Humidifier
On or Off
Waveforms: flow, pressure, volume
Loops: P/V, F/V
all parameters, 72 hours
6 seconds
6 parameters
Audio Level
0 – 120 cmH2O
0 – 120 cmH2O
0 – 120 cmH2O
0 – 120 cmH2O
0 – 3000 ml
0 – 3000 ml