Exterior surface
Clean exterior surfaces by using a soft cloth damped with water and soap. Disinfect exterior
surfaces of the device using isopropyl alcohol.
Clean and disinfect the exterior surfaces of the ventilator after each patient or as required.
Warning
Do not sterilize the ventilator.
Patient tubing system
The patient tubing system components are single-use, dispose of them according to relevant
regulations.
Humidifier
Use the single patient use chamber.
Clean and disinfect the humidifier according to manufacturer instructions. For more
information, please refer to the humidifier manual.
Warning
The humidifier could cause high resistance of the airway and pressure drop.
Check resistance of the airway regularly.
Reusable Expiratory Flow Sensor
Expiratory flow sensor must be cleaned and disinfected according to manufacturer
instructions for each patient or as required.
Singel use Expiratory Flow Sensor
These type of sensors are single use. Dispose of them according to relevant regulations.
Warning
Do not autoclave the expiratory flow sensor.
Clean the expiratory flow sensor immediately after the nebulizer is used.
Nebulizer failure is not detected by a ventilator.