0, NO ERROR DETECTED
There are no error messages in the error buffer
3, OVER TEMPERATURE SHUTDOWN
The temperature inside the controller has exceeded 85
automatically stops any motion that is in progress and changes the status LED color to solid RED.
When the temperature goes below 75
GREEN. Motion related commands can once again be issued in this state.
6, COMMAND DOES NOT EXIST
The command sent to the controller is invalid. Please check the command syntax.
7, PARAMETER OUT OF RANGE
The parameter sent along with a command is out of range.
9, AXIS NUMBER OUT OF RANGE
The axis number specified is out of range.
10, EEPROM WRITE FAILED
Unable to save the controller's settings in its non-volatile memory.
11, EEPROM READ FAILED
Unable to restore the controller's settings from its non-volatile memory.
37, AXIS NUMBER MISSING
The command sent to the controller requires an axis number prefix. Please check the command
syntax.
38, COMMAND PARAMETER MISSING
The command sent to the controller requires a valid parameter. Please check the command syntax.