SC?
RS-485 controller address map query.
Description
SC?
Syntax
This command is used to query the list of all controllers on an RS-485 network.
Remarks
Response
Value
Bit#
0
0
1
1
31
31
Bits 1—31 are one-to-one mapped to controller addresses 1—31. The bit value is set to 1 only
when there are no conflicts with that address.
For example,
1,2, and 7 and more than one controller at address 23, this query will return 135. The binary
representation of 135 is 10000111. Bit #0 = 1 implies that the scan found at least one address
conflict during last scan. Bit #1,2, 7 = 1 implies that the scan found controllers with addresses
1,2, and 7 that do not conflict with any other controller.
SC?
Example
135
SA, SC, SD?
See Also
Value
Integer
Value
Description
0
The scan process did not find any address conflicts
1
The scan process found at least one address conflicts
0
There is no controller with address 1 on the network
1
There is a controller with address 1 on the network
0
There is no controller with address 31 on the network
1
There is a controller with address 31 on the network
if the master controller determines that there are unique controllers at addresses
(get controller address map)
(reply; see above description)
Description
32-bit value. Default = 0
107