4.10

Calibrating Water Temperature

Water temperature is optional.
Water temperature is already calibrated before shipment, so additional calibration is
unnecessary. The calibration procedure is given as a reference. Only when an abnormality
occurs, check that the calibration value is not far from the manually-analyzed value.
4.10.1 Adjustment calibration using sample water <water temperature>
MEASUREMENT screen
See "4.3 Displaying the Calibration Screen" (page 23).
Preparation
Thermometer...1
Calibration procedure
1. Measure water temperature of the sample water using a thermometer other than
this device's and calculate the value.
2. Press the SAMPLE VALUE button ([xx.x]) on the TEMP. CAL. screen.
The sample water calibration value input screen is displayed.
3. Enter the measured temperature of the sample water using the numerical key pad
and press [SET].
Input range: 0.0C to 50.0C
Tip
[CLEAR]: Clears the whole value.
[BACK]: Deletes the rightmost number of the value.
The screen returns to the TEMP. CAL. screen.
4. Press [SAMPLE].
A confirmation screen is displayed asking whether to start calibration.
[CAL.]
CALIBRATION screen
[xx.x]
[TEMP.]
TEMP. CAL. screen
4 CALIBRATION
73