Setting The Calibration Interval; Setting The Standard Time - horiba TW-150 Instruction Manual

Water distribution monitor
Table of Contents

Advertisement

5.3.2

Setting the calibration interval

[MAINTE.]
MAINTENANCE screen
MEAS. screen
See "  Displaying CALIBRATION SETTING screen" (page 82).
1. Press the CAL. INTERVAL button on the CALIBRATION SETTING screen.
The CAL. INTERVAL SETTING screen is displayed.
2. Enter the zero-calibration interval value using the numerical key pad and press
[SET].
Input range: 5 to 9999 (hours)
Tip
[CLEAR]: Clears the entire value.
[BACK]: Deletes the right-most number of the value.
The screen returns to the CALIBRATION SETTING screen.
5.3.3

Setting the STANDARD TIME

[MAINTE.]
MAINTENANCE screen
MEAS. screen
See "  Displaying CALIBRATION SETTING screen" (page 82).
1. Press the STANDARD TIME button on the CALIBRATION SETTING screen.
The STANDARD TIME screen is displayed.
2. Press the button to be changed, and set the value with [
standard time of the ZERO CAL., and press [SET].
The screen returns to the CALIBRATION SETTING screen.
[SETTING]
[CALIBRATION SETTING]
SETTING1/2 screen
CAL. INTERVAL
[xxxxhours]
[SETTING]
[CALIBRATION SETTING]
SETTING1/2 screen
STANDARD TIME
[(Year/Month/day...)]
5 FUNCTION
[CALIBRATION SETTING]
screen
[CALIBRATION SETTING]
screen
] and [
]. Enter the
83