4 CALIBRATION
4.7.3
Adjustment calibration using sample water <conductivity>
MEASUREMENT screen
See "4.3 Displaying the Calibration Screen" (page 23).
Preparation
Conductivity meter...1
Calibration procedure
1. Measure electric conductivity of the prepared sample water using a conductivity
meter other than the device's and calculate the value.
2. Press the SAMPLE VALUE button ([xx.x]) on the COND. CAL. screen.
The sample water calibration value input screen is displayed.
3. Enter the measured value of the sample water using the numerical key pad and
press [SET].
Input range: 0 µS/cm to 1000 µS/cm
Tip
[CLEAR]: Clears the whole value.
[BACK]: Deletes the rightmost number of the value.
The screen returns to the COND. CAL. screen.
4. Press [SAMPLE].
A confirmation screen is displayed asking whether to start calibration.
60
[CAL.]
CALIBRATION screen
[xx.x]
[COND.]
COND. CAL. screen