4.7.2
Individual zero calibration <conductivity>
This is an open zero calibration. Perform calibration only when an abnormality occurs.
MEASUREMENT screen
See "4.3 Displaying the Calibration Screen" (page 23).
Calibration procedure
1. Close the stop valve.
2. Start the water discharge action operation.
Reference
See "5.11 Action" (page 105).
3. Turn off the power.
4. Open the inner door of the device.
Reference
See "2.1 Opening the Inner Door of the Device" (page 7).
[CAL.]
CALIBRATION screen
Stop valve must be closed.
[COND.]
COND. CAL. screen