5.12.4 Analog input check
MEAS. screen
See " Displaying the maintenance item screen" (page 77).
1. Press [ANALOG INPUT] on the CHECK screen.
The ANALOG INPUT screen is displayed.
The value of the analog input from each sensor is displayed.
5.12.5 Analog output check
MEAS. screen
See " Displaying the maintenance item screen" (page 77).
Tip
When there is no operation for 30 minutes ANALOG OUTPUT screen, the display returns to the
previous CHECK screen.
1. Press [ANALOG OUTPUT] on the CHECK screen.
The ANALOG OUTPUT screen is displayed.
2. Press the button of the item to output ([x.xxmA]), and change the values with [
or [
], then press [OUTPUT].
Output value: 0.00/4.00/8.00/12.00/16.00/20.00 mA
3. Check the output with an ammeter.
Reference
"6.1 Analog Output" (page 127)