Calibrating Water Pressure; Adjustment Calibration Using Sample Water <Water Pressure - horiba TW-150 Instruction Manual

Water distribution monitor
Table of Contents

Advertisement

4.9

Calibrating Water Pressure

Water pressure is already calibrated before shipment, so additional calibration is unnecessary.
The calibration procedure is given as a reference. Only when an abnormality occurs, check
that the calibration coefficient is not far from the manually-analyzed value.
4.9.1
Adjustment calibration using sample water <water pressure>
MEASUREMENT screen
See "4.3 Displaying the Calibration Screen" (page 23).
Preparation
Water pressure meter...1
Calibration procedure
1. Measure water pressure of the sample water using a water pressure meter other
than this device's and calculate the value.
2. Press the SAMPLE VALUE button ([x.xxx]) on the PRES. CAL. screen.
The sample water calibration value input screen is displayed.
3. Enter the manually-analyzed value of the sample water using the numerical key pad
and press [SET].
Input range: 0.000 MPa to 1.000 MPa
Tip
[CLEAR]: Clears the whole value.
[BACK]: Deletes the rightmost number of the value.
The screen returns to the PRES. CAL. screen.
4. Press [SAMPLE].
A confirmation screen is displayed asking whether to start calibration.
[CAL.]
CALIBRATION screen
[x.xxx]
[PRES.]
PRES. CAL. screen
4 CALIBRATION
71