5.6

Water Sampling

This function stores sample water when a measurement malfunction occurs. During the set
water sampling time, sample water is obtained from the abnormal water sampling inlet with a
flow rate of 80 mL/min. It is necessary to connect a container to obtain sample water from the
inlet.
Only the first malfunction is obtained, and the time is recorded. Once sample water has been
obtained, the water sampling function is locked until reset. Sample water will not be obtained
even if another malfunction occurs.
The following settings and functions are available for this feature.
Setting
Setting the SAMPLING CHECK
(5.6.1, page 92)
Setting the SAMPLING TIME
(5.6.2, page 92)
Perform the SAMPLING RESET
(5.6.3, page 93)
Displaying the SAMPLING SETTING screen
MEAS. screen
See "  Displaying the maintenance item screen" (page 77).
1. Press [SAMPLING] on the SETTING 1/2 screen.
The SAMPLING SETTING screen is displayed.
The example of connection of the container to obtain a water sampling
Set the timing to obtain the sampling.
Set the time to obtain the sampling.
Obtains sample water once and releases the lock.
When a reset is performed, "LAST SAMPLING TIME"
is cleared from the display.
[MAINTE.]
MAINTENANCE screen
[SAMPLING]
Explanation
[SETTING]
SETTING1/2 screen
ERROR SAMPLE
OUTPUT
Tube
Container for sample water
5 FUNCTION
Initial Setting
OFF
10 minutes
-
91