horiba TW-150 Instruction Manual page 81

Water distribution monitor
Table of Contents

Advertisement

4. Enter the span calibration value using the numerical key pad and press [SET].
Input range: 0 µS/cm to 1000 µS/cm
Tip
[CLEAR]: Clears the whole value.
[BACK]: Deletes the rightmost number of the value.
The screen returns to the COND. CAL. screen.
5. Open the valve of the calibration tank.
6. Press [SPAN].
A confirmation screen is displayed asking whether to start calibration.
7. Press [YES] to start span calibration.
The screen shows that the calibration is in progress.
8. Adjust the flow rate to 50 mL/min using the valve.
Mark showing the 80 mL/min position
50 mL/min position (three gradations below 80 mL/min position)
Flow rate reading position (top surface of the float)
Valve
Wait until the calibration is completed (10 minutes).
Tip
[FIX]: Forcibly stops the calibration and updates it without waiting for automatic completion
[STOP]: Stops the calibration
4 CALIBRATION
55