4. Press the SPAN VALUE button ([xx.x]) on the COL. CAL. screen.
The span calibration value input screen is displayed.
5. Enter the span calibration value using the numerical key pad and press [SET].
Input range: 0.0 TCU to 20.0 TCU
Tip
[CLEAR]: Clears the whole value.
[BACK]: Deletes the rightmost number of the value.
The screen returns to the COL. CAL. screen.
6. Open the valve of the calibration tank.
7. Press [SPAN].
A confirmation screen is displayed asking whether to start calibration.
8. Press [YES] to start span calibration.
The screen shows that the calibration is in progress.
[xx.x]
4 CALIBRATION
35