Common Zero Calibration And Span Calibration <Color - horiba TW-150 Instruction Manual

Water distribution monitor
Table of Contents

Advertisement

4 CALIBRATION
4.5.1
Common zero calibration and span calibration <color>
MEASUREMENT screen
See "4.3 Displaying the Calibration Screen" (page 23).
Preparation
Rubber plug (included)
Calibration tank...1
Span calibration solution...1 L (Pour the solution into the calibration tank)
Used fluid tank (Install it so that used fluid goes into the tank)
Prepare span calibration solution according to the measurement range (0 TCU to 10 TCU or 0
TCU to 20 TCU).
Tip
Select 0 TCU to 10 TCU or 0 TCU to 20 TCU for the measuring range. See "5.10.2 Setting the range for
turbidity and color" (page 101).
Calibration procedure
Perform calibration in the order: common zero calibration → span calibration.
32
[CAL.]
CALIBRATION screen
[COL.]
COL. CAL. screen