horiba TW-150 Instruction Manual page 45

Water distribution monitor
Table of Contents

Advertisement

Calibration information
Calibration
Target
Turbidity
Color
Individual
Conductivity
zero-
pH
calibration
Residual chlorine
Turbidity
Color
Common
Residual chlorine
zero-
calibration
Turbidity
Color
Residual chlorine
Automatic
zero
calibration
Turbidity
Color
Span
Residual chlorine
calibration
pH
Residual chlorine
Turbidity
(zero calibration)
Color
(zero calibration)
Residual chlorine
Adjustment
(span calibration)
calibration
Water pressure
using sample
(zero calibration)
water
pH (zero calibration)
Electric conductivity
(span calibration)
Water temperature
(zero calibration)
Flow rate
It is normal zero calibration performed manually using zero
calibration solution for each measurement item.
50 mL/min
Keep zero calibration solution in the calibration tank. Using
button operation, switch the liquid from the sample water to zero
calibration solution and start calibration.
80 mL/min
It is performed manually using common zero calibration solution
for the three items: turbidity, color and residual chlorine. Filtering
sample water using the filter inside the device generates zero
80 mL/min
calibration solution. Button operation for starting calibration
switches the sample water flow from the normal measurement
line to the filter line and starts calibration.
This operation does not use the calibration tank.
It is performed automatically using common zero calibration
solution for the three items: turbidity, color and residual chlorine.
Filtering sample water using the filter inside the device
generates zero calibration solution. Button operation for starting
calibration switches the sample water flow from the normal
80 mL/min
measurement line to the filter line and starts calibration.
The calibration solution tank is unused as well as common zero
calibration.
You are required to set the first calibration date and time and
interval between calibrations.
It is general span calibration performed manually using span
calibration solution for each measurement item. Keep span
50 mL/min
calibration solution in the calibration tank. Button operation for
starting calibration switches the liquid from the sample water to
span calibration solution and starts calibration.
80 mL/min
It is performed to adjust the value measured by the device to
match the value analyzed manually or measured using another
80 mL/min
instrument using running sample water (tap water). Performed
manually.
Calibration solution and the calibration tank are not used.
4 CALIBRATION
Description
19