Common Zero Calibration And Span Calibration < Turbidity - horiba TW-150 Instruction Manual

Water distribution monitor
Table of Contents

Advertisement

4.6.1
Common zero calibration and span calibration < turbidity >
MEASUREMENT screen
See "4.3 Displaying the Calibration Screen" (page 23).
Preparation
Calibration tank...1
Span standard solution...1 L (Pour the solution into the calibration tank)
Used fluid tank (Install it so that used fluid goes into the tank)
Note
Turbidity span calibration solution tends to settle. Shake thoroughly before using.
Prepare span calibration solution according to the measurement range (0 NTU to 2 NTU, 0
NTU to 5 NTU, or 0 NTU to 10 NTU).
Reference
"5.10.2 Setting the range for turbidity and color" (page 101).
Calibration procedure
Perform calibration in the order: common zero calibration → span calibration.
[CAL.]
CALIBRATION screen
[TURB.]
TURB. CAL. screen