horiba TW-150 Instruction Manual page 54

Water distribution monitor
Table of Contents

Advertisement

4 CALIBRATION
8. Adjust the flow rate to 50 mL/min using the valve.
Wait until the calibration is completed (10 minutes).
Tip
[FIX]: Forcibly stops the calibration and updates it without waiting for automatic completion
[STOP]: Stops the calibration
The screen shows that the calibration is complete.
9. Press [YES].
The pH span calibration coefficient is updated.
10. Close the valve of the calibration tank.
11. Open the stop valve.
12. Adjust the flow rate of the sample water to 80 mL/min using the valve.
13. Disconnect the calibration tank from the calibration inlet of the device.