Individual Zero Calibration And Span Calibration <Color - horiba TW-150 Instruction Manual

Water distribution monitor
Table of Contents

Advertisement

4.5.2
Individual zero calibration and span calibration <color>
MEASUREMENT screen
See "4.3 Displaying the Calibration Screen" (page 23).
Preparation
Calibration tank...2
Zero calibration solution...1 L (Pour the solution into one of the calibration tanks)
Span calibration solution...1 L (Pour the solution into the other calibration tank)
Used fluid tank (Install it so that used fluid goes into the tank)
Calibration procedure
Perform calibration in the order: individual zero calibration → span calibration.
Individual zero calibration
1. Place the calibration tank with zero liquid beside the device.
2. Connect the tank to the calibration inlet of the device.
Calibration
solution tank
3. Press the ZERO VALUE button ([x.x]) on the COL. CAL. screen.
The zero-calibration value input screen is displayed.
[CAL.]
CALIBRATION screen
Connect
[x.x]
[COL.]
COL. CAL. screen
Stop valve
must be closed.
4 CALIBRATION
37

Advertisement

Table of Contents
loading

Table of Contents