horiba TW-150 Instruction Manual page 190

Water distribution monitor
Table of Contents

Advertisement

9 REFERENCE
Installation
environment
This standard is defined in Japanese drinking water test method as of 2017.
No.
Measurement item
1
Turbidity
2
Color
Free residual
3
chlorine (FRC)
4
Water pressure
5
pH
Conductivity
6
(optional)
Water temperature
7
(optional)
*1,*2: Refer to page 162 for correct measurement.
164
Flat and stable location with minimum vibrations or shocks
No dust, mist or corrosive gas in the air
Under atmospheric pressure
No direct sunlight
Good ventilation
Altitude 2000 m or lower
Table 1 Water quality standard
No.
Measurement item
1
Turbidity
2
Color
3
pH
Table 2 Measurement item specifications
Measurement
Measurement range
system
90 degree light
0 to 2 NTU
scattering
0 to 5 NTU
method
0 to 10 NTU
Transmitted
light
0 to 10 TCU
absorbance
0 to 20 TCU
method
Polarographic
0 to 5 mg/L
method
Semiconductor
0 to 1 MPa
detection
0 to 10 Bar
method
Glass electrode
pH 2 to pH 12
method
0 to 500 µS/cm
AC 2-Electrode
0 to 1000 µS/cm
Thermistor
0 to 50C
method
Acceptable range
2 degree or less
5 degree or less
pH 5.8 to pH 8.6
Repeatability
±2.5% of full scale
±5.0% of full scale
±2.5% of full scale
±1.0% of full scale
±0.1 pH
±2.0% of full scale
±0.5C
Calibration method
Formazine standard
1
solution*
2
Standard color solution*
DPD colorimetric method
Standard pressure gauge
pH standard solution
KCl standard solution
Standard thermometer