horiba TW-150 Instruction Manual page 65

Water distribution monitor
Table of Contents

Advertisement

The screen shows that the calibration is complete.
9. Press [YES].
The color zero calibration coefficient is updated.
10. Close the valve of the calibration tank.
11. Disconnect the calibration tank from the calibration inlet of the device.
Span calibration - color
Follow the instructions given in "  Span calibration <color>" (page 34).
4 CALIBRATION