5.14.8 Setting the UNIT HIGH ALARM

[MAINTE.]
MEAS. screen
See "  Displaying the setting screens for each measurement item" (page 99).
1. For each item to be set, press the button of the item on its MEAS. SETTING screen.
Item to be set on the MEAS. SETTING screen:
TURB., COL., FRC., PRES., pH, COND., TEMP.
The MEAS. SETTING screen of each measurement item is displayed.
2. Press the UNIT HIGH ALARM button.
The UNIT HIGH ALARM screen is displayed for the appropriate measurement item.
* This example is for the UNIT HIGH ALARM SETTING screen of TURB.
3. Enter the value using the numerical key pad and press [SET].
Measurement item
Turbidity
Color
Residual chlorine
Water pressure
pH
Conductivity (optional)
Water temperature (optional) 10.0 to 100.0 (C)
Tip
[CLEAR]: Clears the whole value.
[BACK]: Deletes the rightmost number of the value.
The screen returns to the MEAS. SETTING screen for each measurement item.
[SETTING]
MAINTENANCE screen
UNIT HIGH ALARM
[x.xx NTU]
Input range (same for each setting value)
1.00 to 15.00 (NTU)
5.0 to 30.0 (TCU)
1.00 to 7.50 (mg/L)
0.250 to 1.500 (MPa)
2.50 to 15.00 (Bar)
2.00 to 12.00
100 to 1500 (µS/cm)
[I/O]
MEAS.SETTING screen
5 FUNCTION
[(Measurement item)]
MEAS.SETTING screen
Default value
5.00 (NTU)
30.0 (TCU)
3.00 (mg/L)
1.500 (MPa)
12.00
750 (µS/cm)
100.0 (C)
125