Setting The Analog Output Level; Setting Displaying Minus - horiba TW-150 Instruction Manual

Water distribution monitor
Table of Contents

Advertisement

5.2.3

Setting the analog output level

[MAINTE.]
MAINTENANCE screen
MEAS. screen
See "  The I/O SETTING screen" (page 78).
1. Press the ANALOG OUTPUT LEVEL button on the I/O SETTING 2/2 screen.
The ANALOG OUTPUT LEVEL screen is displayed.
2. Select the [0-16] or [4-20] button, and press [SET].
Returns to the I/O SETTING screen.
5.2.4

Setting DISPLAYING MINUS

[MAINTE.]
MAINTENANCE screen
MEAS. screen
See "  The I/O SETTING screen" (page 78).
1. Press the DISPLAYING MINUS button on the I/O SETTING 2/2 screen.
The DISPLAYING MINUS screen is displayed.
2. Select the [ON] or [OFF] button, and press [SET].
Returns to the I/O SETTING screen.
Tip
When changing the setting while the measurement value is being held, the new setting is available
after the HOLD is released.
[SETTING]
SETTING1/2 screen
ANALOG OUTPUT
LEVEL
[0-16mA]
/[4-20mA]
[SETTING]
SETTING1/2 screen
DISPLAYING MINUS
[ON]/[OFF]
5 FUNCTION
[I/O]
I/O SETTING 1/2 screen
[ ]
I/O SETTING 2/2 screen
[I/O]
I/O SETTING 1/2 screen
I/O SETTING 2/2 screen
[ ]
81