Input/Output; Analog Output - horiba TW-150 Instruction Manual

Water distribution monitor
Table of Contents

Advertisement

6

INPUT/OUTPUT

This device has interfaces for parallel input/output, analog output and RS-232C input/output
on the top of the device.
Use the I/O interfaces when you need to output information like warnings and measurement
values, as well as when you need to perform remote operation.
6.1

Analog Output

When outputting to an external device, this device uses an analog current for output.
Measurement items and analog output range, 4 mA to 20 mA
Measurement item
Turbidity
Color
Residual chlorine
Water pressure
pH
Conductivity
(optional)
Water temperature
(optional)
Measurement items and analog output range, 0 mA to 16 mA
Measurement item
Turbidity
Color
Residual chlorine
Water pressure
pH
Conductivity
(optional)
Water temperature
(optional)
Range
0 to 2 NTU
0 to 5 NTU
0 to 10 NTU
0 to 10 TCU
0 o 20 TCU
0 to 2 mg/L
0 to 5 mg/L
0 to 1 MPa
0 to 10 Bar
pH 2 to pH 12
0 to 500 µS/cm
0 to 1000 µS/cm
0to 50C
Range
0 to 2 NTU
0 to 5 NTU
0 to 10 NTU
0 to 10 TCU
0 to 20 TCU
0 to 2 mg/L
0 to 5 mg/L
0 to 1 MPa
0 to 10 Bar
pH 2 to pH 12
0 to 500 µS/cm
0 to 1000 µS/cm
0 to 50C
4 mA output
20 mA output
0.00 NTU
2.00 NTU
0.00 NTU
5.00 NTU
0.00 NTU
10.00 NTU
0.0 TCU
10.0 TCU
0.0 TCU
20.0 TCU
0.00 mg/L
2.00 mg/L
0.00 mg/L
5.00 mg/L
0.000 MPa
1.000 MPa
0 Bar
10 Bar
pH 2.00
pH 12.00
0 µS/cm
500 µS/cm
0 µS/cm
1000 µS/cm
0.0C
50.0C
0 mA output
16 mA output
0.00 NTU
2.00 NTU
0.00 NTU
5.00 NTU
0.00 NTU
10.00 NTU
0.0 TCU
10.0TCU
0.0 TCU
20.0 TCU
0.00 mg/L
2.00 mg/L
0.00 mg/L
5.00 mg/L
0.000 MPa
1.000 MPa
0 Bar
10 Bar
pH 2.00
pH 12.00
0 µS/cm
500 µS/cm
0 µS/cm
1000 µS/cm
0.0C
50.0C
6 INPUT/OUTPUT
127