5 FUNCTION

5.10.3 Correcting color using turbidity

This device can automatically correct the effect of turbidity on the color. However, the degree
of the effect depends on the calibration solution and sample water, and the measurement
value sometimes shifts with sample water. The turbidity coefficient corrects the measurement
value.
[MAINTE.]
MEAS. screen
See "  Displaying the setting screens for each measurement item" (page 99).
Reference
See "  Evaluating the color turbidity coefficient" (page 166).
1. Press the TURB. COEFFICIENTS button on the COL. MEAS. SETTING 1/2 screen.
The COL. TURB. COEFFICIENTS screen is displayed.
2. Enter the COEFFICIENTS value using the numerical key pad and press [SET].
Input range: 0.000 to 9.999  
Tip
[CLEAR]: Clears the whole value.
[BACK]: Deletes the rightmost number of the value.
Returns to the COL. MEAS. SETTING screen.
102
[MEAS. SETTING]
MAINTENANCE screen
TURB. COEFFICIENTS
[Color]
MEAS.SETTING
screen
[x.xxx]
COL. MEAS. SETTING
screen