Checklist; Support Unit - Zeiss KSK 150 FC Instructions For Use Manual

Table of Contents

Advertisement

Operation

Checklist

Support unit

Page 52
For safety reasons, the device must not be used if a malfunction was noticed
before the beginning of an examination or treatment.
• Correct the malfunction (see "What to do in the event of malfunctions")
or contact Carl Zeiss Service.
• Always check the function of the system before examination or treatment
(without patient!) using the following checklist.
 Check that all clamping screws have been firmly tightened.
 Check that the suspension arm has been properly balanced.
 Check that the friction of the rotational motion of the suspension arm, the
carrier arm and the device head has been set correctly.
 Check that the inclination of the device head has been set.
Illumination
 Check that the correct rated voltage has been set for the suspension sys-
tem.
 Check that the power cord is connected.
 Check that the device can be switched on at the power switch (carrier
arm) and the indicator lamp is lit.
 Check that the halogen lamp is intact.
 Check that the halogen illumination can be set to brighter or darker light.
Floor stand
 Check that the locking tabs have been pressed down.
 Check that the stand is securely positioned.
KSK 150 FC
Version 7.1
G-30-1158-en