Operation
Checklist
Support unit
Page 52
For safety reasons, the device must not be used if a malfunction was noticed
before the beginning of an examination or treatment.
• Correct the malfunction (see "What to do in the event of malfunctions")
or contact Carl Zeiss Service.
• Always check the function of the system before examination or treatment
(without patient!) using the following checklist.
Check that all clamping screws have been firmly tightened.
Check that the suspension arm has been properly balanced.
Check that the friction of the rotational motion of the suspension arm, the
carrier arm and the device head has been set correctly.
Check that the inclination of the device head has been set.
Illumination
Check that the correct rated voltage has been set for the suspension sys-
tem.
Check that the power cord is connected.
Check that the device can be switched on at the power switch (carrier
arm) and the indicator lamp is lit.
Check that the halogen lamp is intact.
Check that the halogen illumination can be set to brighter or darker light.
Floor stand
Check that the locking tabs have been pressed down.
Check that the stand is securely positioned.
KSK 150 FC
Version 7.1
G-30-1158-en