System 182
Electrical data
DIQ/S 182-MOD/24V
(24 V AD/DC version)
Electrical connections
DIQ/S 182-MOD
(line power version)
Electrical connections
DIQ/S 182-MOD/24V
(24 V AC/DC version)
ba75504e07
04/2012
Supply
Power consumption
Terminal strip inside the enclosure:
X17
X16
L1
N
100...
240V AC
NETZ/MAINS
Terminal strip inside the enclosure:
X17
X16
24V AC DC
EINGANG
INPUT
POWER
Relay
Output
(3 x)
Max. switching voltage
Max. switching current
Installation
requirements
Relay functions
Nominal voltage:
AC frequency:
Connection:
Line cross-section of connections:
Fuse rating on the operator side: max. 16 A
max. approx. 12 W
X15
X14
X13
X12
X11
X10
£
£
£
240V AC
240V AC
240V AC
£
£
£
2A AC
2A AC
2A AC
R2
R3
R1
X15
X13
X14
X12
X11
X10
£
£
£
240V AC
240V AC
240V AC
£
£
£
2A AC
2A AC
2A AC
R3
R2
R1
Galvanically separated
240 VAC or 24 VDC
2 A (AC and DC)
Fuse rating on the operator side: max. 2 A
Programmable as:
– Opener or closer
– Limit monitor
– Monitoring of the warning and error
signals of the DIQ/S 182-MOD
– Proportional frequency output
– Proportional pulse width output
24 V AC/DC ± 10 %
protective low voltage
SELV (Safety
Extra Low Voltage)
50/60 Hz according to
DIN IEC 60038
2 pin
Europe: 1.5 ... 4.0 mm
USA:
AWG 14 ... 12
X9
X8
X6
X5
X4
HILFS-
SPANNUNG
AUXILIARY
VOLTAGE
SENSORNET 2
AUXILIARY
X9
X8
X6
X5
X4
HILFS-
SPANNUNG
AUXILIARY
VOLTAGE
SENSORNET 2
AUXILIARY
Technical data
2
X3
X2
X1
SENSORNET 1
X3
X2
X1
SENSORNET 1
10 - 3