wtw DIQ/S 182-MOD System Operating Manual page 134

Modular measuring system for 2 digital sensors with modbus rtu/rs 485 output
Table of Contents

Advertisement

Technical data
Mechanical
construction
Ambient conditions
Electrical data
DIQ/S 182-MOD
(line power version)
10 - 2
Enclosure material
Weight
Type of protection
Temperature
Operation
Storage
Relative humidity
Yearly average
Dew formation
Site altitude
Power supply
Protective class
Overvoltage
category
Power consumption
Polycarbonate with 20 % glass fiber
Approx. 0.7 kg
– IP 66
– Corresponds to NEMA 4X
The DIQ/S 182-MOD Universal
Transmitter is not suitable for Conduit
Connection
- 20 °C ... + 55 °C (-4 ... 131 °F)
- 25 °C ... + 65 °C (-13 ... 149 °F)
≤ 90 %
Possible
Max. 2000 m above sea level
Nominal voltage:
100 ... 240 VAC ± 10 %
Frequency:
50/60 Hz
according to
DIN IEC 60038
Line power connection :2 pin, N and L
Line cross-section of mains connection:
Europe: 1.5 ... 4.0 mm
USA:
Fuse rating on the operator side: max. 16 A
II
II
max. approx. 12 W
System 182
2
AWG 14 ... 12
ba75504e07
04/2012