wtw DIQ/S 182-MOD System Operating Manual page 123

Modular measuring system for 2 digital sensors with modbus rtu/rs 485 output
Table of Contents

Advertisement

System 182
ba75504e07
04/2012
Byte
Value
Information
12
42h
Contents of register 13
(HI)
= main measured
value
13
9Eh
Contents of register 13
(LO)
= main measured
value
14
46h
Contents of register 14
(HI)
= main measured
value
15
C2h
Contents of register 14
(LO)
= main measured
value
16
00h
Contents of register 15
(HI)
= secondary measured
value
17
00h
Contents of register 15
(LO)
= secondary measured
value
18
00h
Contents of register 16
(HI)
= secondary measured
value
19
00h
Contents of register 16
(LO)
= secondary measured
value
20
23h
CRC (HI)
21
5Eh
CRC (LO)
Modbus response
Meaning of the value
429E46C2h (Float 32)
--> 79,1382
Measured parameter and
unit,
see byte 10 (measuring
mode)
00000000h (Float 32)
--> 0
but measured value
invalid (MISSING),
see Byte 11 (measured
value status)
Checksum (CRC)