Connecting the voltage supply
Note
The two following paragraphs describe how to connect both models of
the DIQ/S 182-MOD Universal Transmitter to the voltage supply. How
to connect additional power supply modules is described in the
operating manual of the respective power supply module.
3.8.1 DIQ/S 182-MOD (line power version)
Warning
If the power supply is incorrectly connected, it may represent a
danger to life from electric shock. Pay attention to the following
points during installation:
The DIQ/S 182-MOD Universal Transmitter may only be
connected by a trained electrician.
The connection of the DIQ/S 182-MOD Universal Transmitter to
the power supply may only be carried out when it is not
carrying any voltage.
The power supply must fulfill the specifications given on the
nameplate and in chapter 10 T
When installed in a building, a switch or power switch must be
provided as an interrupt facility for the System 182.
The interrupt facility must
– be installed in the vicinity of the DIQ/S 182-MOD Universal
Transmitter, easily accessible by the user, and
– be labeled as the interrupt facility for the
DIQ/S 182-MOD Universal Transmitter.
After the DIQ/S 182-MOD Universal Transmitter has been
installed, it may only be opened if the line voltage has been
switched off beforehand.
Wire end sleeves, suitable for the power line, with suitable crimping
tool
1 x screwed cable gland with sealing ring (scope of delivery of the
Universal Transmitter).
Cable stripping knife
Tools
Wire stripper
Phillips screwdriver
Small screwdriver.
.
ECHNICAL DATA
Installation
3 - 29