wtw DIQ/S 182-MOD System Operating Manual page 90

Modular measuring system for 2 digital sensors with modbus rtu/rs 485 output
Table of Contents

Advertisement

Relay outputs
Monitoring limiting
values using one or two
relays
6 - 6
Measured
t1
value
Relay 1
UL
2
1
LL
Relay 2
Fig. 6-1
Switching points for relays with the function of a limit indicator
1
Upper limit value (relay 1) exceeded
2
Selected switching delay t1 for relay 1 expired
Relay 1 switches
3
Hysteresis for upper limiting value (relay 1) undercut
4
Selected switching delay t1 for relay 1 expired
Relay 1 switches back
5
Lower limit value (relay 2) undercut
Selected switching delay t2 for relay 2 expired
Relay 2 switches
6
Hysteresis for lower limiting value (relay 2) exceeded
Selected switching delay t2 for relay 2 expired
Relay 2 switches back
A switching delay (t) can be set up for each relay for switching
processes. This is the time period for which a limiting value must be
exceeded before the relay switches. This prevents frequent switching if
the measured values are close to the limiting value.
t1
4
3
5
t2
System 182
Hysteresis UL
6
Hysteresis LL
t2
ba75504e07
Time
04/2012