wtw DIQ/S 182-MOD System Operating Manual page 76

Modular measuring system for 2 digital sensors with modbus rtu/rs 485 output
Table of Contents

Advertisement

Operation
Putting the sensor into
operation after
calibrating
4 - 16
5
Confirm the calibration data with
with this. The following display message describes the further
steps to put the sensor into operation again.
6
Confirm with
display (the measured value flashes as the sensor is still in the
maintenance condition).
7
If calibration was successful, immerse the sensor in the test
sample.
8
Wait for a stable measured value.
9
Switch off the maintenance condition (highlight sensor and
g
press
three times). The measured value has stopped
flashing.
Note
If calibration was not successful, this is indicated by "----" on the display
and a corresponding message with remedial actions appears in the log
book. Follow the instructions and repeat calibration.
g
g
. The display returns to the measured value
System 182
. Calibration is completed
ba75504e07
04/2012