7.14.3 High-level settings

Different reactions can be chosen when the level sensor detects the height of the heat carrier liquid
level. Depending on the setup, heat carrier liquid or operation conditions, one of the following settings
may be suitable:
Setting
No warning
Warning
Warning and
heater off
Alarm
Master
LMnet
4x
zu
RDC--
zu
lGVW
Command
Over level handling none
Help
Menu
94
Master
Command
settings
settings
lGlml
none
lGVW
Warning
lGVWG
Warning +
heater off
lG@K@
Alarm
lGVW
− Choose the menu for the high-level reaction.
− The factory setting is
2x
− Press
°C
− Choose a setting from the above list.
Warning
Warn.+ Heater off
Alarm
End
T
set
Proline Low-temperature Thermostats
Reaction and application recommendation
Select only when no safety sensitive application. I.e.
water as heat carrier liquid.
Acoustic and optical warning as long as the level goes
down. This is the factory setting.
Warning and additional heater off as long as the level
goes down. Recommended for flammable heat carrier
liquids with much higher flash point and temperatures
above 100°C.
Alarm switches off the pump and the heater until the
alarm is removed by pressing
keyboard. Recommended for external loads and
flammable liquids.
lGVW
to enter the setup mode. The current setting flashes.
Over level handling
− Open the device parameter menu via the
soft key
− With the cursor keys continue to
Settings
− The shown window appears
− Select the preferred parameter with
− See introduction for details.
T
fix
Starting up
on the Master
. That is a warning.
Menu
.
Over level handling
and confirm with
.
YACE0072 / 21.08.07
.
or