Restoring The Works Setting Of The Internal Temperature-Probe Offset - Lauda Proline RP 845 C Operating Instructions Manual

Table of Contents

Advertisement

Temperature value of the ref.
temp. measurement device:
Min: -50.0°C
1
Help
6
Menu

7.8.8 Restoring the works setting of the internal temperature-probe offset

If the offset has been misadjusted unintentionally, the works setting can be restored with this function.
Master
MNEnu
3x
to
CAL..
2x
to
1x
dEF 1
3 sec
Command
intern Pt100
extern Pt100
Pump
Menu
54
20.
15
Max: 302.00
2
3
4
7
Screen
8
T
9
set
dEF 1
 Call offset works setting for the internal temperature probe
1
dEF 1
°C
 The works setting is accepted after pressing for 3 seconds and
donE
Calibration
Default
End
T
set
Proline Edition X cooling thermostats
 The temperature measurement device
 Change the display in the adjacent
 or quit the window with
5
T
0
fix
.
is displayed.
is displayed.
Default
 Open the device parameter menu via
 With the cursor keys continue to
 The window shown adjacent appears.
 Confirm selection with
T
fix
Starting up
shows the true temperature value (with
glass thermometers consider the
correction where applicable!).
window to the true value with cursor or
soft keys and accept with
changes.
the soft key
Menu .
Settings
Calibration
intern Pt100
Default
YACE0100 / October 4, 2018
or End ,
without making
dEF
.
.