7.6.7 Restoring the works setting of the internal temperature-probe offset

If the offset has been misadjusted unintentionally, the works setting can be restored with this function.
Command
intern Pt100
extern Pt100
Help
Menu
Confirm input!
Enter key: Continue
Escape key: Cancel
Help
Menu
56
20.
15
Max: 302.00
2
3
4
7
Screen
8
T
9
set
Calibration
Default
End
T
set
End
T
set
− The temperature measurement device
shows the true temperature value (with
glass thermometers consider the
correction where applicable!).
− Change the display in the adjacent
window to the true value with cursor or
soft keys and accept with
− or quit the window with
changes.
5
T
0
fix
Default
− Open the device parameter menu via
the soft key
− With the cursor keys continue to
intern Pt100
− The window shown adjacent appears.
− Confirm selection with
T
fix
− Confirm the control dialog on the right
with
− Return to the measurement window with
End or
T
fix
Starting up
Proline Kryomats
Menu .
Settings
Calibration
Default
.
or cancel with
.
YAUE0007 / 2/26/2009
or End ,
without making
.
.