7.8.8 Restoring the works setting of the internal temperature-probe offset

If the offset has been misadjusted unintentionally, the works setting can be restored with this function.
Master
3x
to
CAL..
2x
to
1x
dEF 1
3 seconds
Command
intern Pt100
extern Pt100
Pump
Menu
x
YACE0099 / 10/9/2018
20,
15
Max: 302,00
2
3
Menu
7
Screen
8
T
dEF 1
 Call offset works setting for the internal temperature probe
MNEnu
°C
 The works setting is accepted after pressing for 3 seconds and
Calibration
Default
End
T
4
5
9
T
0
set
fix
dEF 1
,
dEF 1
is displayed.
donE
is displayed.
T
set
f
Starting up
 The temperature measurement device
shows the true temperature value (with
glass thermometers consider the correc-
tion where applicable!).
 Change the display in the adjacent win-
dow to the true value with cursor or soft
keys and accept with
 or quit the window with
changes.
Default
 Open the device parameter menu via
the soft key
Menu .
 With the cursor keys continue to:
Settings
Calibration
intern Pt100
Default
 The window shown adjacent appears.
 Confirm selection with
or End ,
without making
.
.
55