Defining temperature limits

With this function, it is possible to define a minimum and a maximum temperature in which the
thermostat controls. By reaching the temperature limits, a warning appears. In this way setpoint input
can be prevented which may damage the heat transfer liquid or the apparatus. For example, if water
were used as the heat transfer liquid, +95 °C would be practicable as the maximum temperature and
+5 °C as the minimum temperature.