7.6.2
Defining temperature limits
With this function, it is possible to define a minimum and a maximum temperature in which the
thermostat controls. By reaching the temperature limits, a warning appears. In this way setpoint input
can be prevented which may damage the heat transfer liquid or the apparatus. For example, if water
were used as the heat transfer liquid, +95 °C would be practicable as the maximum temperature and
+5 °C as the minimum temperature.
Command
Pump
Settings
Graph
Clock
Programmer
Interfaces
Control
Limits
Pump
Menu
Lower limit (T il)
Min: -100.0 °C
1
Help
6
Menu
62
T il (min) -50.0 °C
T ih (max) 90.0 °C
End
T
set
-50.
0
Max: 202.0 °C
2
3
4
7
Screen
8
T
9
set
Limits
Open the device parameter menu via the
soft key
With the cursor keys continue to
The minimum and maximum
temperatures are displayed.
T il (min)
Select the limit to be changed with
T
fix
Enter the desired limit temperature.
Accept the change with
or quit the window with
changes.
5
T
0
fix
Starting up
Proline Kryomats
Menu .
is currently active.
and confirm with
.
without making
YAUE0007 / 10/2/2018
Limits
.
or
,