Activating The "Standby" Operating State; Defining Temperature Limits - Lauda Viscotemp 15 G Operation Manual

Table of Contents

Advertisement

7.4.4

Activating the "Standby" operating state

In the "Standby" mode the pump, heater and chiller are switched off. The operating display remains active.
7.4.5

Defining temperature limits

With this function the temperature limits Til and Tih are defined. If, for example, you are using water as the heat transfer
liquid, +5 °C is practicable as the minimum temperature and +95 °C as the maximum temperature.
By pressing
-
09/2021
The adjacent menu window appears.
In the entry window the cursor flashes below the value to be changed.
The permissible adjustment range is indicated with Min and Max.
(ESC) you are returned to the menu level without any change.
Activate Standby by pressing
-
Access to the main menu level is obtained by pressing the enter
-
key
.
Selection and confirmation of 
-
Select the lower (Til) or upper (Tih) limit with
-
firm it with
.
Change the value with
-
Single figures can be selected by pressing
-
By pressing
(+/-) the arithmetic sign can be changed.
-
Confirm your choice with the enter key
-
ECO Gold
(right soft key).
Setup
 Limits.
or
or
.
or
.
.
and con-
51 / 134