7.6
Defining temperature limits
With this function, it is possible to define a minimum and a maximum temperature in which the thermostat controls. By
reaching the temperature limits, a warning appears. In this way setpoint input can be prevented which may damage the
heat transfer medium or the apparatus. For example, if water is used as the heat transfer liquid, 95 °C would be practi-
cable as the maximum temperature and 5 °C as the minimum temperature. The default values are Hi = 105 °C and
Low = -30 °C.
2x for
3x for
and 1x
H i
L o
or
Wait 4 seconds or
7.7
Refrigeration unit submenu
Menu
only with cooling thermostats.
Cool
4x
to
and 1x
or
Wait 4 seconds or
28 / 52
Hi
Select option
Hi
or option
Lo
Enter the selection
The maximum temperature can be set from 35 °C to 105 °C.
The minimum temperature can be set from -30 °C to 30 °C.
The current temperature limit value is displayed flashing.
Set the required temperature limit.
Display flashes 4 seconds new value is automatically accepted, or
By pressing the <Enter> key the value is accepted immediately.
CooL
Call the option
CooL
Configure the refrigeration system with these possible settings:
Auto
on
oFF
The selected mode is displayed flashing.
Display flashes 4 s new selection is automatically accepted,
or
it is immediately accepted with the <Enter> key.
Lo
and
Hi
(maximum temperature),
Lo
(minimum temperature) using the key
CooL
.
automatic refrigeration mode;
refrigeration on duty;
refrigeration out of duty.
Alpha
.
7/2022