7.5.9
Limiting the mains current consumption
If your mains fusing is below 16 A, the current consumption can be reduced in steps from 16 A to 8 A.
The maximum heating power is then also reduced correspondingly. Here, take into consideration
whether other loads are connected to the same fused circuit or whether your ECO thermostat is the only
load.
Basic setup
Acoustic signal
Display
Autostart
Resolution
Curr.Consumpt. 16.0A
DLK connected
Language
סּOK
ESC
Curr.Consumpt.
Max:16,0
Min:8,0
16,00
סּOK
ESC
44
−
Access to the main menu level is obtained by pressing
►
►
−
Selection and confirmation of
on►
The adjacent menu window appears.
−
As well as
►
−
Select
Standby
−
Change the value with
−
By pressing
vidually and change them with
−
You confirm the set value by pressing
−
You quit the window without changes with
Standby
Putting the device into operation
.
Curr.Consumpt.
the set value is displayed.
Curr.Consumpt.
with
or
or
you can also select numbers indi-
ECO GOLD
Setup
Basic
setup.
.
.
or
.
.
(ESC).