Technical Data

12.2 Test Data

12.2
Test Data

Reference Conditions according to IEC 62586-1 for Determining Test Data

12.2.1
Ambient temperature
Relative humidity
Supply voltage
Phases (3-wire network)
External continuous magnetic fields
DC components V/I
Signal waveform
Frequency
Voltage magnitude
Flicker
Unbalance (all channels)
Harmonic
Interharmonic

Electrical Tests

12.2.2
Standards
Standards
Insulation Test according to IEC EN 61010-1 and IEC EN 61010-2-030
Inputs/Outputs
Current measurement
inputs
Voltage measurement
inputs
Supply voltage
Binary outputs
Ethernet interface
RS485 interface
EMC Tests for Immunity (Type Tests)
Standards
Electrostatic discharge,
Class III, IEC 61000-4-2
250
Insulation
Rated Voltage
Reinforced
150 V
Reinforced
480 V
Reinforced
300 V
Reinforced
300 V
Function
< 50 V
Function
< 50 V
23 °C ± 2 °C
40 % to 60 % RH
V
± 1 %
PS
3
DC field: ≤ 40 A/m
AC field: ≤ 3 A/m
None
Sinus
50 Hz ± 0.5 Hz
60 Hz ± 0.5 Hz
Udin ± 1 %
Pst < 0.1 %
100 % ± 0.5 % of Udin
0 % to 3 % of Udin
0 % to 0.5 % of Udin
IEC EN 61000-6-2
IEC EN 61000-6-4
IEC EN 61010-1
IEC EN 61010-2-030
ISO Test Voltage
AC 2.3 kV
Surge voltage
9.76 kV
DC 3.125 kV
AC 3.536 kV
DC 700 V
DC 700 V
IEC EN 61000-6-2
For more standards see also individual functions
6 kV contact discharge
8 kV air discharge, both polarities
150 pF, Ri = 330 Ω with connected Ethernet cable
SICAM, SICAM P850/P855 7KG85X, Manual
E50417-H1040-C482-A9, Edition 03.2022
Category
Cat. III
Cat. III
Cat. III
Cat. III