Technical data
Operable flow range
Output signal
Signal on alarm
96
Maximum full scale value for gas [kg/h]
g
max(G)
g
Maximum full scale value for liquid [kg/h]
max(F)
< g
g
g
max(G)
max(F)
r
Gas density in [kg/m
G
DN
[mm]
8
15
25
40
50
To calculate the measuring range, use the Applicator sizing tool (® ä 108)
Calculation example for gas
• Sensor: Promass E, DN 50
• Gas: Air with a density of 60.3 kg/m
• Measuring range (liquid): 70 000 kg/h
3
• x = 125 kg/m
(for Promass E DN 50)
Maximum possible full scale value:
× r
: x = 70 000 kg/h × 60.3 kg/m
g
= g
max(G)
max(F)
G
Recommended measuring range
"Flow limit" section (® ä 101)
Greater than 1000 : 1. Flow rates above the preset full scale value do not overload the amplifier so
the totalized values are registered correctly.
17.4

Output

2-wire
• Current output 1: 4-20 mA HART, passive
• Current output 2: 4-20 mA, passive
• Lower-range value, full scale value and time constant (0.07 to 999 s) can be set
Depending on the interface, failure information is displayed as follows:
• Current output
– Failsafe mode selectable (in accordance with NAMUR Recommendation NE 43):
Minimum alarm: 3.6 mA
Maximum alarm (= factory setting): 22 mA
– Failsafe mode with user-selectable value: 3.59 to 22.5 mA
• Local display
– Status signal (in accordance with NAMUR Recommendation NE 107)
– Plain text display
• Operating tool via HART communication or service interface (CDI)
– Status signal (in accordance with NAMUR Recommendation NE 107)
– Plain text display
can never be greater than g
max(G)
3
] at operating conditions
[in]
3/8
½
1
1 ½
2
3
(at 20 °C and 50 bar)
3
Proline Promass E 200 HART
max(F)
x
[kg/m
3
]
85
110
125
125
125
: 125 kg/m
3
= 33 800 kg/h
Endress+Hauser