Diagnostics and troubleshooting
82
Diagnostic event and event text
The fault can be identified using the diagnostic event. The event text helps you do so by providing
information about the fault. In addition, the corresponding symbol for event level is displayed before
the diagnostic event.
Symbol for event
level
Example
If two or more diagnostic messages are pending simultaneously, only the message with the highest
priority is shown. Additional pending diagnostic messages can be shown in the Diagnostics list
submenu (® ä 83).
Past diagnostic messages that are no longer pending are shown in the Event logbook submenu
(® ä 87).
Operating elements
Operating functions in menu, submenu
Plus key
Opens the message about the remedial measures.
A0013970
Enter key
Opens the operating menu.
A0013952
12.2.2

Calling up remedial measures

X X X X X X X
X
X
20.50
1
Event text
2
Symbol for event level with diagnostic event
3
Event ID for service
4
Operation time of occurrence
5
Remedial measures
The user is in the diagnostic message.
Diagnostic event
Status signal
¯
A0013962
A0013958
3-digit number
Message for remedial measures
S
X X X X X X X
Curr. output 1
mA
Curr. output 1
1
2
1. Check process
2. Check current output settings
5
Proline Promass E 200 HART
Event number
Event text
¯
¯
441
Curr.output 1
X
X
S441
i
(ID:153)
S441
2d11h05m55s
S
3
4
A0013940-EN
Endress+Hauser