Endress+Hauser HART Proline Promass E 200 Operating Instructions Manual page 99

Coriolis mass flow measuring system
Hide thumbs Also See for HART Proline Promass E 200:
Table of Contents

Advertisement

Proline Promass E 200 HART
Repeatability
Response time
Endress+Hauser
Example for max. measured error
[%]
±2.5
±2.0
±1.5
±1.0
±0.5
0
0
1
2
å 19
Maximum measured error in % o.r. (example: DN 25)
Design fundamentals (® ä 100)
Flow values (example)
Turn down
250 : 1
100 : 1
25 : 1
10 : 1
2 : 1
3
o.r. = of reading; 1 g/cm
= 1 kg/l; T = fluid temperature
Mass flow and volume flow (liquids)
±0.125 % o.r.
Mass flow (gases)
±0.35 % o.r.
Design fundamentals (® ä 100)
Density (liquids)
±0.00025 g/cm
3
Temperature
±0.25 °C ± 0.0025 × T °C (±0.45 °F ± 0.0025 × (T–32) °F)
• The response time depends on the configuration (damping).
• Response time in the event of erratic changes in the measured variable: after 500 ms 95 % of the
full scale value
4
6
8
10
Flow
[kg/h]
72
180
720
1 800
9 000
Technical data
12
14
16
18
Maximum measured
error
[lb/min]
[% o.r.]
2.646
6.615
26.46
66.15
0.25
330.75
0.25
t/h
A0015463
3
1.2
0.3
99