Mounting
22
6.2.3

Mounting the measuring device

WARNING
Danger due to insufficient process sealing.
Ensure that the inside diameter of the seals is greater than or equal to that of the measuring tube
and piping.
Ensure that the seals are clean and undamaged.
Fasten the seals correctly.
1. Ensure that the direction of the arrow on the nameplate of the sensor matches the flow direction
of the fluid.
2. Install the measuring device or turn the transmitter housing so that the cable entries do not
point upwards.
Ã
6.2.4

Turning the transmitter housing

To provide easier access to the connection compartment or display module, the transmitter housing
can be turned:
8 mm
1. Unscrew the securing screw using an open-ended wrench.
2. Rotate the housing in the desired direction.
3. Firmly tighten the securing screw.
6.2.5

Turning the display module

Proline Promass E 200 HART
max. 350°
8 mm
1
3 mm
Endress+Hauser
A0013964
A0013713
A0013905