8 Technical data
Measured variable
Measuring range
Output signal
Galvanic isolation
Electrical connections
Supply voltage
Power consumption
Reference operating con-
ditions
Max. deviation at the out-
puts
Max. measured error at the
inputs
Ambient temperature
Electromagnetic
compatibility (EMC)
24
8
Technical data
8.1
Technical data at a glance
8.1.1
Input
Current input, frequency input
0 ... 25 mA, 0 ...15000 Hz
8.1.2
Output
Electromagnetic flowmeter: Simulation voltage 0.15 ... 20 mV
yes
8.1.3
Power supply
Input for power unit: 100...240 V AC (47-63 Hz) Prim., 12 V DC (2.5 A) Sec.
12 V DC, internal battery 7.2 V, 3800 mAh
140 mA
Charging current: max. 2A (max. 350 mA for simultaneous operation)
8.1.4
Performance characteristics
+22 °C ± 2K
Electromagnetic measuring device
±0.2 % of the full scale value voltage (0.150 ... 20 mV)
Current
± 5 µ A (0 ... 25 mA)
Frequency
< 0.5 Hz (0 ...15 kHz)
8.1.5
Operating conditions (Environment)
0 ... 60 °C
In accordance with EN 61326/A1 and NAMUR recommendation NE 21
PROline FieldCheck™
Endress+Hauser