Proline Promass E 200 HART
Endress+Hauser
Parameter overview with brief description
Parameters
Prerequisite
Decimal places 1
A measured value is
specified in the
parameter Value 1
display.
Decimal places 2
A measured value is
specified in the
parameter Value 2
display.
Decimal places 3
A measured value is
specified in the
parameter Value 3
display.
Decimal places 4
A measured value is
specified in the
parameter Value 4
display.
Display interval
-
Header
-
Header text
The Free text option
is selected in the
Header parameter.
Separator
-
10.6

Configuration management

After commissioning, you can save the current device configuration, copy it to another measuring
point or restore the previous device configuration. You can do so using the Configuration
management parameter and its options.
Navigation path
Setup ® Advanced setup ® Conf.backup disp ® Config. managem.
Functions of the parameter options
Options
Description
Execute backup
A backup copy of the current device configuration in the HistoROM is saved to the display
module of the device. The backup copy comprises the transmitter data of the device.
Restore
The last backup copy of the device configuration is copied from the display module to the
HistoROM of the device. The backup copy comprises the transmitter data of the device.
Duplicate
The transmitter configuration from another device is duplicated to the device using the
display module.
Description
Selection/
User entry
Select the number of
• x
digits after the decimal
• x.x
point for the display
• x.xx
value.
• x.xxx
• x.xxxx
Select the number of
• x
digits after the decimal
• x.x
point for the display
• x.xx
value.
• x.xxx
• x.xxxx
Select the number of
• x
digits after the decimal
• x.x
point for the display
• x.xx
value.
• x.xxx
• x.xxxx
Select the number of
• x
digits after the decimal
• x.x
point for the display
• x.xx
value.
• x.xxx
• x.xxxx
Set the length of time
1 to 10 s
the measured values
are displayed if the
values alternate on the
display.
Select the content for
• Device tag
the header of the local
• Free text
display.
Enter the text for the
Max. 12 characters
header of the local
such as letters,
display.
numbers or special
characters
Select the decimal
• . (point)
separator for
• , (comma)
numerical values.
Commissioning
Factory setting
x.xx
x.xx
x.xx
x.xx
5 s
Device tag
- - - - - - - - - - - -
. (point)
71