Proline Promass E 200 HART
Endress+Hauser
1. Press the O key (
symbol).
à The message for the remedial measures for the diagnostic event opens.
2. Press S + O simultaneously.
à The message about the remedial measures closes.
12.3
Diagnostic event in the operating tool
If a diagnostic event is present in the operating tool, the status signal appears in the top left status
area along with the corresponding symbol for event level in accordance with NAMUR NE 107:
• Failure (F)
• Function check (C)
• Out of specification (S)
• Maintenance required (M)
Calling up remedial measures
1. Navigate to the "Diagnostics" menu.
à In the "Actual diagnostics" parameter, the diagnostic event is shown with event text.
2. On the right in the display range, hover the cursor over the "Actual diagnostics" parameter.
à A tool tip with remedial measures for the diagnostic event appears.
12.4
Diagnostic list
In the Diagnostics list submenu, up to 5 currently pending diagnostic messages can be displayed.
If more than 5 messages are pending, the messages with the highest priority are shown on the display.
Navigation path
Diagnostics ® Diagnostics list
Diagnostics and troubleshooting
A0014014-EN
83