Technical data
Cable entries
Cable specification
Reference operating
conditions
Maximum measured error
98
• Cable gland: M20 × 1.5 with cable Æ 6 to 12 mm (0.24 to 0.47 in)
• Thread for cable entry: ½" NPT, G ½"
• Permitted temperature range: –40 °C (–40 °F) to ³ 80 °C (176 °F); at least ambient temperature
+ 20 K
• A normal device cable suffices if only the analog signal is used.
• A shielded cable is recommended if using the HART protocol. Observe grounding concept of the
plant.
17.6

Performance characteristics

• Error limits following ISO/DIN 11631
• Water with +15 to +45 °C (+59 to +113 °F) at 2 to 6 bar (29 to 87 psi)
• Specifications as per calibration protocol ±5 °C (±9 °F) and ±2 bar (±29 psi)
• Accuracy based on accredited calibration rigs that are traced to ISO 17025
To obtain measured errors, use the Applicator sizing tool (® ä 108)
In addition to the values indicated, the measured error at the current output is typically ±4 mA.
o.r. = of reading; 1 g/cm
3
= 1 kg/l; T = fluid temperature
Mass flow and volume flow (liquids)
±0.25 % o.r.
Mass flow (gases)
±0.75 % o.r.
Design fundamentals (® ä 100)
Density (liquids)
• Reference conditions: ±0.0005 g/cm
• Field density calibration: ±0.0005 g/cm
(valid after a field density calibration under process conditions)
• Standard density calibration: ±0.02 g/cm
(valid over the entire temperature range and density range (® ä 101))
Temperature
±0.5 °C ± 0.005 × T °C (±0.9 °F ± 0.005 × (T – 32) °F)
Zero point stability
DN
[mm]
8
15
25
40
50
3
3
3
[in]
[kg/h]
3/8
0.24
½
0.78
1
2.16
5.40
2
8.40
Proline Promass E 200 HART
Zero point stability
[lb/min]
0.0088
0.0287
0.0794
0.1985
0.3087